BP 3AG1 Blood pressure monitor Model no.: BP 3AG1. Dimensions: 140 x 120 x 70 mm. Weight: 400g (including batteries) Voltage source: 4 x 1.5 V alkaline batteries; size AA or mains adapter DC 6V, 600 mA (optional)
For most Microlife blood pressure monitors, the error 3 indicates that the unit cannot reach the proper pressure in the cuff to start the test. If replacing the batteries doesn't work, check that the cuff is properly plugged into the socket on the monitor. Next you may need to replace the cuff. There may be a leak in the cuff or the hose between the cuff and the socket. Microlife sells replacement blood pressure cuffs here: https://www.microlife.com/consumer-products/blood-pressure/cuffs-and-supplies . These cuffs fit all of their devices except for the wrist versions.
Repeat measurement keeping arm still. If inflation of the cuff takes too long, the cuff is not correctly seated or the hose connection's not tight. Check connections and repeat
ERROR 2 IN MICROLIFE PRESSURE MONITOR MEANS : Unnatural pressure impulses influence the measurement result. Reason: The arm was moved during the measurement (Artefact).
×